DescriptionSTD-TE 60 is a three-station reciprocating tribometer, designed around a high pressure c
Aplikasi WhatsApp: +86 13816217984
Email :info@qinsun-lab.com
STD-TE 60 is a three-station reciprocating tribometer, designed around a high pressure chamber, for experiments under hydrogen and other gases. It includes a vacuum pump for purging the chamber before charging with a chosen gas, so can also be used for tests under medium vacuum. Tests may be run with plate specimen temperatures from -50°C to +150°C.
Technical Specifications | |
Number of Test Stations: | Three |
Test Configuration: | Ball on Flat |
Cylinder on Flat | |
Gases: | Hydrogen, Argon, CO2 |
Working Pressure: | 0.1 mbar to 100 bar |
Proof Pressure: | 120 bar |
Stroke Range: | 0 to 20 mm Continuously Adjustable |
Frequency: | 0 to 5 Hz |
Motion: | Sinusoidal |
Loading System: | Tension Spring |
Load Range: | 5 to 100 N |
Friction Range: | 0 to 50 N |
Deflection Sensor: | Inductive Transducer |
Pressure Rating: | 150 bar |
Refrigerant Probe: | |
Action: | Contact cooling of plate specimen |
Temperature Range: | Ambient to -50°C |
Heating Probe: | |
Action: | Contact heating of plate specimen |
Temperature Range: | Ambient to 150°C |
Electrical Feedthroughs: | PT100 x 2 |
Rotary Feedthough: | BMD-300 Magnetic Coupling |
Motor: | NX-210 a.c. Servo Motor |
Automatically Controlled Parameters | |
Frequency | |
Plate specimen temperature | |
Test duration | |
Manually Controlled Parameters | |
Load | |
Stroke | |
Chamber pressure (external service) | |
Measured Parameters | |
Friction | |
Plate specimen temperature | |
Chamber temperature | |
Chamber pressure | |
Number of cycles | |
Derived Parameters | |
Friction coefficient | |
Services | |
Electricity: | 220/240V, single phase 50/60 Hz, 7.5 kW |
Process Gas System: | Not included |
ATex Rating | The machine, drive motor and associated instrumentation are ATex rated. The chiller unit, vacuum pump, power module and PC control and data acquisition system are not ATex rated. It follows that these items must be installed outside any ATex rated test cell, with suitable ATex rated connections between the operator work-station and the test cell. |
Penafian: Konten di atas hanya untuk referensi dan komunikasi antara orang dalam industri, dan tidak menjamin akurasi atau kelengkapannya. Menurut hukum dan peraturan yang relevan dan peraturan situs web ini, unit atau individu yang membeli barang terkait harus memperoleh kualifikasi dan kondisi kualifikasi yang valid.