Product Introduction:YY/T 0810 and ASTM F1800:The standards specify the test methods for evaluating
Email: info@standard-groups.com
Product Introduction:
YY/T 0810 and ASTM F1800:
The standards specify the test methods for evaluating the fatigue performance of tibial trays in knee prostheses, which support and protect the plastic articular surfaces. These methods apply to tibial trays that cover both the medial and lateral tibial condyles. The test involves cantilever loading at a frequency not exceeding 10 Hz, with an R-ratio of 0.1, and is conducted for 5 million cycles or until the specimen fails.
Kale Control's M electromagnetic series equipment offers precise sinusoidal load control, providing comprehensive test fixtures that meet the experimental conditions required by the standards. Additionally, we deliver accurate and effective test protocols (SOPs) to assist in the mechanical performance analysis of tibial trays.
Penafian: Konten di atas hanya untuk referensi dan komunikasi antara orang dalam industri, dan tidak menjamin akurasi atau kelengkapannya. Menurut hukum dan peraturan yang relevan dan peraturan situs web ini, unit atau individu yang membeli barang terkait harus memperoleh kualifikasi dan kondisi kualifikasi yang valid.