Total Hip Joint Prostheses Wear Tester - Qinsun
Manufacturer Instrumen Pengujian Profesional
Pantas konsultasi nasional:+86 021-64208466

Total Hip Joint Prostheses Wear Tester

- Evaluating the wear - resistance of hip joint prostheses: By simulating human movement states such

WhatsApp: +86 13816217984

Email: info@qinsun-lab.com

Konten

Standard 

  • YY/T 0651.1-2016 (China): Specifies wear testing parameters for total hip prostheses, including angular motion, loading modes, and environmental conditions.
  • ISO 14242-1:2014 (International): Defines protocols for simulating physiological loading and kinematics during hip joint wear testing.
  • ASTM F3047M-15: Additional compliance for advanced testing scenarios including edge loading and micro-separation.

Product Advantages

  1. Triaxial Motion Simulation: Replicates full physiological motion profiles (flexion/extension ±100°, abduction/adduction ±30°, internal/external rotation ±40°).
  2. Dynamic Load Control: Dual-peak loading up to 4,500 N with 0.1% force accuracy.
  3. Biorelevant Environment: Temperature-controlled (37±2°C) lubrication bath with 30±2 g/L protein concentration.
  4. Extended Durability: Capable of 5+ million cycle testing with <0.5% drift in calibration.
  5. Automated Compliance: Pre-programmed ISO/YYT test protocols with real-time data validation.
  6. Modular Design: Six-station configuration allows parallel testing of multiple specimens.

Product Features

  • Electrohydraulic Actuation: Six independent servo-hydraulic actuators for precise motion control.
  • 3D Force/Torque Sensors: Multi-axis sensors measure axial forces and torsional moments simultaneously.
  • Temperature-Controlled Bath: Recirculating lubrication system with protein concentration monitoring.
  • High-Resolution Imaging: Optional integration with micro-CT systems for in-situ wear analysis.
  • AI-Powered Analytics: Machine learning algorithms predict wear rates from initial test cycles.
  • Safety System: Emergency stop, over-pressure relief, and laser-based specimen breakage detection.

Parameter Teknik

ParameterSpecification
Angular Motion RangeFlexion/Extension: ±100°
Abduction/Adduction: ±30°
Rotation: ±40°
Max. Axial Force±4,500 N (1,000 N resolution)
Load Control Accuracy±0.5% of reading (0–4,500 N)
Frequency Range0.1–5 Hz (programmable waveforms)
Temperature Control37±2°C (PT1000 sensors)
Lubrication System15L capacity with protein skimmer and bubble trap
Data Acquisition100 kHz sampling rate (16-bit resolution)
Power Requirements220V/110V ±10%, 50/60Hz, 2,500W
Dimensions (W×D×H)1,800 × 1,200 × 2,000 mm
Berat950 kg (including bath and control cabinet)
Noise Level<65 dB (A-weighted)

Standard Accessories

  1. Hip Simulator Fixture Kit:
    • Self-centering femoral head adapters (16–32 mm)
    • Acetabular cup positioning jigs (60° inclination standard)
  2. Biorelevant Bath System:
    • 15L capacity with recirculation pump and protein skimmer
    • 316L stainless steel construction with HEPA filtration
  3. WearMaster Pro Software:
    • Automated wear rate calculation (mass loss/volume displacement)
    • 21 CFR Part 11-compliant audit trail with blockchain timestamping
    • Remote monitoring via iOS/Android app
  4. Operator Kit:
    • Calibrated torque wrench set (1–50 Nm)
    • Specimen alignment gauges (ISO 14242-1 compliant)
  5. Calibration Bundle:
    • NIST-traceable force calibration weights
    • Angular encoder verification tool
  6. Safety Kit:
    • Laser safety curtains
    • Acoustic enclosure with 30 dB noise reduction


Penafian: Konten di atas hanya untuk referensi dan komunikasi antara orang dalam industri, dan tidak menjamin akurasi atau kelengkapannya. Menurut hukum dan peraturan yang relevan dan peraturan situs web ini, unit atau individu yang membeli barang terkait harus memperoleh kualifikasi dan kondisi kualifikasi yang valid.