Guide pin anchoring force tester - Qinsun
Manufacturer Instrumen Pengujian Profesional
Pantas konsultasi nasional:+86 021-64208466

Guide pin anchoring force tester

Product introduction  The anchoring force test of the guide needle can ensure the anchoring effect o

Email: info@standard-groups.com

Konten

Product introduction


  The anchoring force test of the guide needle can ensure the anchoring effect of the mammary 

positioning filament within the tissue and prevent the filament from shifting or falling out before 

surgery. This is essential for stable positioning filaments to accurately locate and remove the tumor 

during surgery.


Test standard


  YY/T0880-2013 "Disposable Breast Positioning wire and Guide Needle"


Technical parameter


  1. Tensile force test range: 0——20N, error is not more than +1%, resolution 0.01N


  2. Test time: 1-999S, any choice, the error is not more than +0.5S


  3. Test speed: 10mm/min, 50mm/min, 100m/min


  4. Distance adjustment: 10mm——50mm adjustable


  5. Sensor: high-precision force value sensor (with hardware and software protection function)


  6. Fixture: high quality 304 stainless steel


  7. Force measuring mechanism: built-in


  8. Printer: on-board micro needle printer


  9. Data storage: USB disk can be imported into the computer storage


  10. Machine power: 70W


  11. Power supply: AC90V-240V/50Hz


导引针锚定力测试仪1.jpg

导引针锚定力测试仪2.jpg




Penafian: Konten di atas hanya untuk referensi dan komunikasi antara orang dalam industri, dan tidak menjamin akurasi atau kelengkapannya. Menurut hukum dan peraturan yang relevan dan peraturan situs web ini, unit atau individu yang membeli barang terkait harus memperoleh kualifikasi dan kondisi kualifikasi yang valid.