Medical mask bacterial filtration efficiency (BFE) testerIt is composed of aerosol generation system
Email: info@standard-groups.com
Medical mask bacterial filtration efficiency (BFE) tester
It is composed of aerosol generation system, aerosol chamber and aerosol transmission device, negative pressure cabinet and 28.3l /min sampler.The whole tester is controlled by the console.The control console adopts microcomputer control system to coordinate and control the work of aerosol generation system, transmission system, negative pressure cabinet and sampling system, and display the working status in real time. The whole measurement work is completed automatically.
Mask Bacterial Filtration Efficiency Tester Features
Key Specification | Specification Range | Resolution | Max. permissible errors |
A Route Sampling Flow | 28.3L/min | 0.1L/min | Within ± 2.0% |
B Route Sampling Flow | 28.3L/min | 0.1L/min | Within ±2.0% |
Spray Flow | (8~10)L/min | 0.1L/min | Within ±5.0% |
Peristaltic Pump Flow | (0.006~3.0)mL/min | 0.001ml/min | Within ±2.0% |
Front Pressure of A Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of B Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.0% |
Front Pressure of Spray Flowmeter | (0~300)kPa | 0.1kPa | Within ±2.0% |
Negative Pressure of the Chamber | (-90~-120)Pa | 0.1Pa | Within ±1.0% |
Working Temperature | 0-50℃ | ||
Negative Pressure of the Chamber | >120Pa | ||
Data Save Ability | >500000 sets(Scalable capacity) |
Penafian: Konten di atas hanya untuk referensi dan komunikasi antara orang dalam industri, dan tidak menjamin akurasi atau kelengkapannya. Menurut hukum dan peraturan yang relevan dan peraturan situs web ini, unit atau individu yang membeli barang terkait harus memperoleh kualifikasi dan kondisi kualifikasi yang valid.