Welding heat tester drops 600°C cylinder to measure material protection.
Email: info@standard-groups.com
The instrument is suitable for determining the protection provided by materials under heat exposure when welding or with other related procedures.
A steel cylinder in a steel cone and in an oven is heated to a temperature of 600°C. After the heating the steel cone is put onto a mechanism which is on top of the calorimeter and the specimen. When triggering the magnetic switch the steel cylinder vertically falls onto the specimen. The calorimeter measures the starting temperature, the maximum temperature and the maximum difference. Before and after the test the calorimeter is adjusted to 20°C by a cooling and heating unit. The software manages the controlling, calculations and evaluation of the standard.
Parameter Teknik
Item | Parameter |
Sample geometry | Textiles |
Test standards | ISO DIS 9150 |
Sensory function | Temperature (calorimeter) |
Kuasa | 230 VAC / 150 VA |
Dimension | 600 x 300 x 300 mm (W x D x H) |
Berat | 20kg |
Penafian: Konten di atas hanya untuk referensi dan komunikasi antara orang dalam industri, dan tidak menjamin akurasi atau kelengkapannya. Menurut hukum dan peraturan yang relevan dan peraturan situs web ini, unit atau individu yang membeli barang terkait harus memperoleh kualifikasi dan kondisi kualifikasi yang valid.